Reports of 78 Clear search Modify search
AdV-TCS (CO2 laser projector)
nardecchia, cesarini, gherardini, menzione - 22:33 Thursday 16 February 2023 (58919) Print this report
NI CO2 actuator alignment attempt with HWS-DET

Today it has been decided to perform the HWS-DET measurement to check the alignment CH vs DAS.

Before starting, Fabio and Nicola restored the same positions for BS/SR/NI/SDB1/SDB2 during the HWS-DET measurement performed 'in gentle unlock configuration' Saturday Feb 11th (fig. 1).
The HWS-DET measurement started at 16.07 UTC, the reference has been acquired with the inner ring switched OFF and outer ring and CH ON at the steady state.
Thus, at around 16.15 UTC, the outer ring has been switched OFF to monitor the cooling phase.
The OPL induced by turning OFF the DAS appeared in the map (see fig. 2).
Then, between the WF 173 and WF 174 acquired at 17.41.12 UTC and 17.41.45 UTC, respectively,  the beam disappeared from the map (see fig.3). Looking at the signals, something happened on SDB1 LC_TX/TY (see fig.4).

The measurement was aborted at 17.51 UTC.

Images attached to this report
Comments to this report:
rocchi - 9:50 Friday 17 February 2023 (58922) Print this report

I looked around to try to understand what happened. This is from the VPM log:

2023-02-16-17h41m31-UTC    info menzione     'Bench:Use floating setpoints [-1:Disable,1:Enable=1]' sent to SDB1_LC

It seems quite in coincidence with the movement of the bench.

Likely, this was an action part of the recovery after the activities on injection/laser (58918).

Images attached to this comment
gouaty, masserot - 17:53 Friday 17 February 2023 (58929) Print this report

We analyzed the incident that occured with SDB1_LC.

Indeed at 17h41m31-UTC the operator requested the engagement of the SDB1 floating setpoints. At that time the bench goes to a position corresponding to the last values memorized in the floating setpoints.

The issue comes from the fact that the memory of the filters has been resetted on Tuesday Feb 14 at 11h24m02-UTC when restarting the process SDB1_LC (following the works mentioned in https://logbook.virgo-gw.eu/virgo/?r=58882). So now the floating setpoints are zeroed, which corresponds to a bad alignment state of the bench.

I have put back the bench in fixed setpoints. We should keep using the fixed setpoints until the B5 beam drift control is successfully closed. Once the B5 beam drift will be closed, then one can request the use of floating setpoints.

Images attached to this comment
AdV-TCS (CO2 laser projector)
cesarini, melo, lorenzini, taranto - 15:32 Saturday 09 July 2022 (56397) Print this report
DAS intensity distribution and alignment check

Yesterday morning we continued the work on CO2 benches.

We started with the WI bench, and we checked the intensity distribution of the two DAS rings, we adjusted a little bit both the Inner and Outer rings, since the outer doesn’t enter in the camera in the image plane, we took 4 different photos (see figure 1 and 2).

We verified the percentages of transmission and reflection of the pickoff power meter and they are 14% and 86% respectively. We checked again the alignment in the far field and near field (see figure 3 and 4), it was still good.

At this point, we moved on the NI bench and we checked the intensity distribution and the alignment after that we rotated the BSrec in order that the three actuators enter completely in the Pickoff as done for the WI bench (see entry 56383).

Also in this bench we did a small movement to adjust the intensity distribution and we had to realign well the DASes on the CH. The final results are shown in figure 5, 6 and 7.

We verified the percentages of transmission and reflection of the pickoff power meter and they are 14% and 86% respectively, as for the WI bench.

At the end, we left the TCS configuration:

 

CH [mW]

DAS IN [mW]

DAS OUT [mW]

WI

120

100

2600

NI

160

350

2400

Images attached to this report
AdV-TCS (CO2 laser projector)
cesarini, nardecchia, lorenzini - 12:04 Sunday 08 March 2020 (48654) Print this report
Power Change on WI CO2 Laser Bench

Today at 8.00 UTC again a thermal state change of the CO2 laser on WI bench happened. In attachment the VIM detail on WI power and temperatures (fig1., fig.2). After a fast check, we can say that the jump down of the WI CO2 power can be linked to an increase of the laser temperature, but it is not related to any change in room temperature or chiller temperature. Furthermore the NI does not show analog behaviour (see fig.3). The ITF state seems to be not affected by this change. We will monitor powers and temperatures and regular visual checks to chillers will be asked to operator. 

Images attached to this report
AdV-TCS (Ring Heater)
nardecchia, cesarini, ruggi, mantovani, tacca - 16:00 Wednesday 08 January 2020 (48097) Print this report
Comment to TCS RH tuning (48090)

At 14.55 UTC we decreased the power applied to the ETM RHs:

P NE: 6.7 W--> 6 W

P WE: 7 W--> 6.3 W 

AdV-TCS (Ring Heater)
cesarini, nardecchia, lumaca, mantovani - 10:20 Wednesday 08 January 2020 (48090) Print this report
TCS RH tuning

In order to rule out the possibility that the 150kHz is due to a mechanical resonance in the long arm cavity we have decreased emitted RH power both on WE and NE. At 9:02 UTC we have set the NE RH voltage to 15.79 V and WE RH voltage to 16.14 V. 

Comments to this report:
ruggi - 15:04 Wednesday 08 January 2020 (48095) Print this report

The action on RH did not solve the problem, but something has changed. First of all, the locks are shorter. There is also a change in the shape of the instability: after the tuning, it is sistematically faster. In fig 1 we can see the last 50 s of the latest lock stretches: the one in the middle is the one during which the tuning has been performed. In fig 2 we can see the death of all the locks since yesterday: the change seems systematic.

Images attached to this comment
nardecchia, cesarini, ruggi, mantovani, tacca - 16:00 Wednesday 08 January 2020 (48097) Print this report

At 14.55 UTC we decreased the power applied to the ETM RHs:

P NE: 6.7 W--> 6 W

P WE: 7 W--> 6.3 W 

nardecchia, tacca, berni - 14:52 Saturday 18 January 2020 (48188) Print this report

I'm going on with the scan of the ETM RHs.

At 13.41 UTC I decreased the power applied on both RHs by 0.5 W, so:

-WE: 6.3 W--> 5.8 W

-NE:5.9 W--> 5.4 W

nardecchia, tacca, magazzu - 19:41 Saturday 18 January 2020 (48189) Print this report

Since a general worsening of the ITF happened after the RH power change (see figure in attachment), it has been decided to go back in two steps. 

As first step, at 18.30 UTC, I set:

-PWE: 5.8 W-->6.1 W

-PNE:  5.4 W--> 5.7 W

Images attached to this comment
AdV-TCS (Ring Heater)
cesarini, nardecchia - 16:47 Thursday 24 October 2019 (47372) Print this report
WE RH tuning

The power applied to the WE RH has been increased by 100 mW at 14.41 UTC.

AdV-COM (AdV commissioning (1st part) )
cesarini, mantovani, lorenzini, rocchi - 16:27 Friday 18 October 2019 (47281) Print this report
summary of the 18th morning shift

Today we continued the tuning of TCS parameters.

We tried to increase power on NI CP to increase CMRF adjusting both NI DAS_OUT and NI DAS_IN. We found at the end a no good status of the entire ITF, both sideband power and CMRF were bad. We are not sure this was up to our work, but in any case we decided to restore the situation obtained on 17th at 14.00 UTC.
New DAS parameters are:
NI DAS_IN :     400 mW
NI DAS_OUT : 2.38 W
WI DAS_IN :    159 mw
WI DAS_OUT: 191 mW
Trend will be shown.

 
AdV-COM (AdV commissioning (1st part) )
mantovani, cesarini, lorenzini - 14:56 Friday 18 October 2019 (47276) Print this report
Summary of the 17th morning shift

Yesterday morning we have started the tuning of the DAS actuators following this procedure:

starting point: CMRF = -0.4e-9; 112MHz mag = 0.081; CD = 19;

- 1st step: increase of the sideband power using the WI outer ring (from 170 mW to 200 mW)

CMRF = -0.5e-9; 112MHz mag = 0.087; CD = 18.4;

- 2nd step: decrease the CMRF using the NI outer ring (from 2.36 W to 2.41 W). This tuining is not differential thus the common mode (the one that maximise the sidebands)  is spoiled  and has to be compensated

CMRF = -0.12e-9; 112MHz mag = 0.083; CD = 18.6;

- 3rd step: increase back the sidebands using the WI inner ring (from 154mW to 124 mW)

CMRF = it was @ LN1.  112MHz mag = 0.085; CD = 18.6;

the trend is visible in Figures 1 and 2  to highilight the pourpouse of the actuator tuning).
Figure 3 shows the tuning to reduce the CMRF.

Images attached to this report
Comments to this report:
mwas - 15:43 Friday 18 October 2019 (47279) Print this report

In addition to the 56MHz sideband gain, the B1p power / contrast defect, and the CMRF, the power on B1s2 when in LN3 is an important figure of merit. The B1s2 power when both OMCs are locked is dominated by the 56MHz TEM01 mode, which causes OMC lock acquisition issues and couples the 56MHz RAM to the sensitivity. From that perspective the best state was at around 9UTC with about 0.075mW, this is better that other times, but still higher than the 0.05mW we had during O3a.

Images attached to this comment
AdV-TCS (CO2 laser projector)
Aiello, Nardecchia, Rocchi, Fafone, Lorenzini, Cesarini - 20:22 Thursday 16 November 2017 (40055) Print this report
CO2 DAS shining tests

This morning we have shined both CPs with DAS in order to observe their effect on the ITF.
We shined both NI and WI DAS with three different power levels:

Start (UTC) End (UTC) P inner DAS (mW) P outer DAS (mW)
10:00 11:29 50 100
11:30 12:40 150 300
12:41 13:45 300 600

As expected from simulations (VIR-0855A-17), we observed a decrease of the BNS horizon and of the gain of 56 MHz sidebands (see attached plot). In particular, the measured gain of the 56 MHz sideband with 0.3 W inner DAS power and 0.6 W outer power is in fair agreement with that computed with SIS (VIR-0889A-17).

Images attached to this report
Comments to this report:
schaaf, tacca - 17:25 Monday 20 November 2017 (40094) Print this report

Phase camera images obtained during the test.

1. figure attached shows the 6 MHz overlap (cleanest and most responsive signal obtained with the phase camera in this test) with the different gpstimes at which actions took place:

  •     red stretches indicate where the DAS power was adjusted
  •     black lines indicate where the images shown in B1p.png and B4.png are obtained
  •     green shows where the modulation index of the 6 MHz sidebands is increased
  •     no black line (and thus images) without DAS turned on in order not to get fooled by the changes due to the increased modulation index


2. figure attached shows the power images on B4

  •     power in sidebands is reduced
  •     astigmatism in sidebands is reduced
  •     56 MHz LSB starts to wander away from the carrier spot location


3. figure attached shows the power images on B1p

  •     misalignment becomes visible in the carrier
  •     80 mode survives
  •     faint outer rings start to appear for the 6 MHz LSB
  •     56 MHz sidebands and with some imagination the 6 MHz USB seem to get squashed in vertical direction


4. figure attached shows the overlaps observed with the phase camera

  •     6 MHz overlaps (with carrier) are (slightly) improved with modulation index increase
  •     Worsening of the overlaps when the power of the DAS is increased but stabilizes after roughly half an hour in all cases (with exception of the final unlock ...)
     
Images attached to this comment
AdV-TCS (CO2 laser projector)
cesarini, fafone, lorenzini, nardecchia, rocchi - 10:02 Thursday 23 March 2017 (36996) Print this report
Comment to WI CH centering (36930)

The wavefront images acquired with the RH on (2017-03-15) allowed to identify the ITM center in the HWS-INJ reference systems (36791, VIR-0097A-17):

  • X=[-1.85+/-0.05]mm
  • Y=[1.25+/-0.04]mm

The wavefront images acquired with the HWS-INJ with CH on (figure 1 shows an example) have been analyzed to check the alignment on the CP center. 

To evaluate the coordinates of the minimum Delta OPL in the CH case we have considered 1.30 h of data:

  • X:[-1.84+/-1.00]mm
  • Y:[ 1.80+/-1.20]mm

The CH is aligned on the CP center within 0.6 cm.

Images attached to this comment
AdV-TCS (CO2 laser projector)
cesarini lorenzini nardecchia rocchi - 20:07 Wednesday 15 March 2017 (36930) Print this report
WI CH centering
This afternoon shift has been dedicated to the centering of the WI CH on the CP.
At 11.43 UTC we started the HWS acquisition.
The WI CH has been turned on at maximum power at 11.51 UTC.
In the meanwhile we have analyzed the wavefronts maps to evaluate the position of the minimum of DeltaOPL caused by CH.
From a preliminary analysis, CH seems to be centered on the CP. The coordinates of the mirror center in the HWS maps have been found using the RH (VIR-0176A-17).
A deeper data analysis will follow.
The WI CH has been turned off at 16.13 UTC.
Images attached to this report
Comments to this report:
cesarini, fafone, lorenzini, nardecchia, rocchi - 10:02 Thursday 23 March 2017 (36996) Print this report

The wavefront images acquired with the RH on (2017-03-15) allowed to identify the ITM center in the HWS-INJ reference systems (36791, VIR-0097A-17):

  • X=[-1.85+/-0.05]mm
  • Y=[1.25+/-0.04]mm

The wavefront images acquired with the HWS-INJ with CH on (figure 1 shows an example) have been analyzed to check the alignment on the CP center. 

To evaluate the coordinates of the minimum Delta OPL in the CH case we have considered 1.30 h of data:

  • X:[-1.84+/-1.00]mm
  • Y:[ 1.80+/-1.20]mm

The CH is aligned on the CP center within 0.6 cm.

Images attached to this comment
AdV-TCS (CO2 laser projector)
cesarini, lorenzini, nocera, paoletti, sposito - 12:28 Wednesday 15 March 2017 (36923) Print this report
Status of the WI CO2 laser

Yesterday morning at the beginning of our shift we noticed that the WI CO2 laser was shining a lower total power (about  25% less). Looking at the data it seems that this strange behaviour has begun on 7/03/2017 afternoon (around 17.00 UTC) after 4 hours of large instability (see fig1). During the shift, the WI laser was switched off to allow actions on NI bench (moving the RF TTL generator closer to the laser RF driver). When we  tried to switch it on again, it did not shine anymore. Since the two diagnostic LEDs on the RF driver where red we supposed a failure of the driver itself. Today we replaced the damaged driver with a new spare and the laser is now working properly.  During the replacement we noticed that the mammut  and the cables that connect the RF driver with the power supply were a little bit burnt (see fig 2 and 3).  We replaced the connector and we separated the power supply on two different cables feeding the two RF modules (before the power supply was connected to the mammut by a single cable). 

Images attached to this report
AdV-TCS (CO2 laser projector)
cesarini, lorenzini - 12:04 Wednesday 15 March 2017 (36922) Print this report
TCS shift - NI CH

Yesterday 14 March we worked on NI bench trying to understand why the beam size was not as expected by our simulations (see 36836). After some trials, we decided to look at the beam without the spatial filtering, removing the pinhole, after and before the CH lens. The spot is not as expected and it seems to be different in shape and dimension from previous images in same conditions (31/1/2017 and 21/02/2017). Further investigations needed.

At the end of the shift we realized that the WI CO2 laser was not shining anymore. A detailed entry will follow.
 

AdV-TCS (CO2 laser projector)
cesarini - 11:59 Tuesday 21 February 2017 (36599) Print this report
Comment to Central heating on WI (36570)

The iris has been checked, the dimension of the hole was 2.65 mm in diameter instead of 2.65 mm in radius. We check both FFT and Zemax simulations, placing the iris at the object plane of the CP,  the right dimension is 2.65 mm of radius. The loss of power, according to simulations, should be about 10% and the shape on the CP circular. See fig.1 for the cross section of the central heating on the CP simulated by FFt and Zemax. The iris has been machined again and will be placed on WI during the maintenance of  the 21 Feb.

Images attached to this comment
AdV-COM (AdV commissioning (1st part) )
fiori, cesarini, paoletti, aiello, lorenzini - 16:52 Friday 17 February 2017 (36569) Print this report
TCS chiller 1Hz noise: solved!

As planned (Federico's suggestion) this morning the settings of the two chillers were carefully compared.  Indeed Elisabetta and Lorenzo found that the (silent) NI chiller had "energy save mode" set on, while the (noisy) WI chiller had not. At 8:35 LT the WI chiller was set in "energy save mode". No more 1Hz noise has been observed since then, see Figure 1. The temperature stability performance of the two chillers look similar (Figure 2).

Images attached to this report
AdV-TCS (CO2 laser projector)
Aiello, Cesarini, Lorenzini - 16:48 Friday 17 February 2017 (36570) Print this report
Central heating on WI
Today we worked on the WI CO2 laser bench to change the central heating lens, in order to increase the dimension of the central heating on the CP. We replaced the previous lens (focal length of 140 mm) with a new one of 50 mm of focal length. We needed to realign the CO2 beam on cross hair. We tried to put an iris in the object plane of the CP to avoid the elliptical shape generated by in vacuum mirror clipping. After the positioning of the iris we checked the power transmitted and we realized that about 50% of the power is clipped, so we decided to remove it. We need to further check simulations to find out the suitable correction of the diameter of the iris.
During the shift, we connected the two CH pick-off power meters to the ADC channels, now two channels TCS_WI_CO2_POWER_CH_PICKOFF and TCS_NI_CO2_POWER_CH_PICKOFF are created that provide directly the expected power on CH, based on the calibration of the ratio between CH power and Pick-off power performed in the past. The maximum available CH power on WI is about 220 mW with a PSD reading of 0.57 W (and has to be proportionally rescaled to different values of PSD reading).

Comments to this report:
cesarini - 11:59 Tuesday 21 February 2017 (36599) Print this report

The iris has been checked, the dimension of the hole was 2.65 mm in diameter instead of 2.65 mm in radius. We check both FFT and Zemax simulations, placing the iris at the object plane of the CP,  the right dimension is 2.65 mm of radius. The loss of power, according to simulations, should be about 10% and the shape on the CP circular. See fig.1 for the cross section of the central heating on the CP simulated by FFt and Zemax. The iris has been machined again and will be placed on WI during the maintenance of  the 21 Feb.

Images attached to this comment
lorenzini, nardecchia - 16:08 Tuesday 21 February 2017 (36605) Print this report
The new iris machined with 2.65 mm radius has been placed this morning on the WI bench, in the CP object plane before the CH lens. The power loss was measured to be about 13%.
Environmental Monitoring (Environmental Monitoring)
fiori, paoletti, cesarini, lorenzini, aiello, carbognani - 14:23 Friday 17 February 2017 (36557) Print this report
TCS chiller 1Hz noise, results from the last tests

In the last week a few tests have been performed with TCS chillers to investigate better and find possible mitigations. Mitigation attempts are up to now not effective. Although some strategies are still to be attempted at the chiller side, we will try to investigate also the the possibility to mitigate the coupling. Here a summary of the latest attempts and possibile further strategies:

  1. on Feb 9 the heater of the WI chiller has been disconnected. Of course the noise disappeared, but in spite of many trials modifying control parameters, the temperature stability could not be recovered. One limitation is that the chiller firmware does not allow to completely disable the heating loop. The WI chiller heater was re-connected on Feb 13 but the WI chiller was kept off till Feb 16.
  2. on Feb 14 the NI chiller has been switched on alone (i.e. while the WI chiller was off). All along this time the noise never appeared and as it was confirmed by the chiller on-board display, the heater never turned on (note that the chiller does not provide remote information on the heating status, the information is provided only at the on-board display). It was also tried to increase the T set point from 19°C  to 22°C: the heater turned just for a few minutes during the transition, than it kept silent again. See Figures 1 (UPS spectrogram 24h Feb 14), 2 (UPS spectrogram 1hour during T set change) and 3 (chiller temperature and T set). We have to conclude that "something is different with the NI chiller" which makes it never turn on the heater. (for conclusions see point 2. below)
  3. on Feb 16 the WI heater was plugged on IPS (ENEL) instead than on UPS as usual, and switched on. We observed that the 1Hz noise at the level of INJ signals somehow reduced (a factor 2 or 3) but it was still present. As expected, the 1Hz comb noise disappeared from UPS and appeared on IPS, the noise is roughly the same at the magnetometer location: Figure 4 (shows noise with WI chiller on IPS compared to chiller OFF) and Figure 5 (compares noise with WI chiller on IPS and on UPS). (for conclusions see point 3 below)

Conclusions and further steps:

  1. operating the chiller with the heater off seems not feasible.
  2. we will further investigate for possible differences between chiller NI (not noisy) and WI (noisy). One possiblity is that there is a difference at the level of the firmware or the settings (indeed the Chiller technician we spoke with mentioned that up to his knowledge the heater PWM cycle is 3Hz, while in order to explain our observations it should be 1Hz cycle) . Another possibility is that the "magic" condition of the NI chiller is due to the difference in thermal load for example because of the less heat dissipation in the shorter water pipes (also to be checked if there is a different regulation of the water flow).
  3. The fact that the noise does not disappear when the chiller is powered on ENEL, makes us suspect  that even plugging it on a dedicated UPS would not solve the problem.
  4. One possible mitigation strategy is to replace the chiller with that of Virgo (which was only cooling) and its custom control (however we know from data that the old chiller performance was at the level of 0.1°C at the most, while the new chillers performs at 0.01°C, and seems that the new CO2 lasers need this level of T stab.)
  5. Another strategy that we will start pursuing is to investigate further the coupling path and act on it.
Comments to this report:
fiori - 16:04 Friday 17 February 2017 (36567) Print this report

missing plots are here

Images attached to this comment
AdV-TCS (CO2 laser projector)
cesarini, aiello, lorenzini, nardecchia - 14:57 Tuesday 14 February 2017 (36507) Print this report
Central heating

During the maintenance  at 8.00 LT we opened the enclosure of the WI CO2 laser bench and set the laser in a pulsed way to allow the positioning of  the pin hole. We profiled again the beam to recover the same beam size. We also put in place  the power meter for the central heating pick off. The last part of the cabling to the ADC is missing.  At 12.50 we tested the laser at full power and everything was ok. At 13.05 we switched off the WI CO2 laser and its chiller.

At 10.20 we switched on the NI CO2 laser and we positioned the power meter for the central heating pick off. The last part of the cabling to ADC is missing. At the end of the maintenance we left the NI CO2 laser switched on, with enclosure closed, to test the 1 Hz magnetic noise difference between two chillers.

 
 

AdV-TCS (CO2 laser projector)
cesarini, lorenzini, nardecchia, paoletti - 18:05 Monday 13 February 2017 (36499) Print this report
WI CO2 laser status

The WI CO2 laser and its chiller have been switched off at 16.40 LT to connect again the cable of the chiller heater, that was disconnected to remove the 1Hz magnetic noise (36441). The laser has been put back in operation at maximum power at 17.45 LT. 

AdV-TCS (CO2 laser projector)
lorenzini, cesarini - 12:50 Monday 23 January 2017 (36159) Print this report
TCS shift
During the shift of 22/1/17 morning we accessed the WI CO2 bench and continue the characterization of the CH beam, in particular we looked at the beam in far field with and without the spatial filtering (that is aimed at optimizing the DAS profile). The analysis of the gathered data is ongoing, by comparison with Zemax and FFT simulations.
AdV-TCS (CO2 laser projector)
cesarini, lorenzini, mantovani, masserot - 10:10 Friday 30 December 2016 (35942) Print this report
CO2 laser slow power control

In order to implement the slow power control on CO2 lasers we tested the communication between DAQ and Pzt driver and we tried to make a scan between 0 and 50 V. This procedure is essentisal to find the control working point  in a region of linearity.

First of all, the TCS_CEB has been modified in order to be able to drive the TCS DAC (ch0 and ch1 of the first TCS DAC). The noise features have been added (line, ramp, white noise, colored noise) with output TCS_NOISE (which shows the output voltage of the Pzt driver). The noise is driven to the two DAC channels thanks to a driving matrix obtaining the two DAC signals (TCS_N_NOISE and TCS_W_NOISE for the North and West CO2 laser benches respectively showing the voltage sent by the DAC).

After that, we changed the gain of our Pzt driver from 7.5 to 10 and we set a ramp on the DAC of 5 V. Due to differential mode of the DAQ and  single ended mode of the Pzt driver we have a problem of floating ground and half of the voltage at the output. So, to recover this problem we set 10 V ramp on the DAC that gives a maximum output of 5V and a maximum voltage at the Pzt driver output of 50 V. We show in figure the TCS_NOISE channel that gives the Pzt driver output, TCS_N_NOISE and   TCS_W_NOISE, that gives the DAC outputs, the photodiodes ISSIN and ISSOUT for both benches, that monitor the laser power.

We can observe a wide region of linearity suitable to find a good working point. Now the control with UFG of 0.1 Hz can be implemented.

Images attached to this report
Comments to this report:
masserot - 10:43 Friday 30 December 2016 (35950) Print this report

Please try to connect correctly the DAC differential  output to the single input of the driver by connecting 

  • the DAC_pin+ to driver+ and the Dac_ground to the driver_ground
  • instead of  the DAC_pin+ to driver+ and the Dac_pin- to the driver_ground as it is according the logbook entry

You may damage the DAC  output stage or the driver input stage

AdV-COM (AdV commissioning (1st part) )
Izumi, Cesarini, Lorenzini - 0:47 Friday 30 December 2016 (35944) Print this report
TCS test with CITF: not convincing

In this afternoon we spent a shift working on the TCS tuning with CITF.

We haven't decided on the working point for the central heating yet as the interpretation of the CITF data is non-trivial.

At this point I might suggest, as a different strategy, relying on the Hartman cameras and providing the dead-reckoned thermal lens values.


[The tests]

We tried two particular CO2 settings as follows.

  • Configuration 1:
    • This is to provide a 20 uD single-path lens on each arm to force the power recycling cavity stable.
    • [CO2 NI, CO2 WI] = [140 mW, 140 mW] . These numbers were derived theoretically.
    • rotation stage settings [theta NI,  theta WI] = [35 deg, -22 deg]
    • Initially locked on the carrier, then switched to the sideband in the middle of the measurement as the carrier lock wasn't able to maintain a stable lock.
  • Configuration 2:
    • This to set the CO2s at the point where we had a highest build up in the past (35867).
    • [CO2 NI, CO2 WI] = [20 mW, 20 mW]
    • rotation stage [theta NI, theta WI] = [-3 deg, -40 deg]
    • Locked on the carrier throughout the measurement.

[Results]

  • Conf 1:
    • Significant degradation in the build up. This was true for both carrier- and sideband-locked CITF. Consistent with the past measurement (35867).
    • B4 image showed Laguerre-type modes both for carrier- and sideband-locked CITF. This is also consistent with the past measurement.
  • Conf 2:
    • No significant change in the DC signals.

In addition to these measurement, the highest build up of 15 mW at B4_DC was observed before these measurements when neither of the CO2s was activated. The B2_DC signal also dropped to 4 mW (from 13.8 mW) which is the lowest value I have seen. This is inconsistent with all the past measurements we did with CITF before. This could mean that there are uncontrolled variables somewhere affecting the build up of CITF. Note that since the arm cavities weren't available today, we set the NI alignment using the B2 and B4 cameras/DC signals as referecnes.

[Current interferometer state]

Both NI and WI central heating are switched off. The CITF locking loops are off. Also the below are the demod phases that we found (kind of) optimum for this afternoon.

  • B1p_119MHz_phi0 = -0.72 rad
  • B2_6MHz_phi0 = -0.67 rad
AdV-TCS (Wavefront sensing)
cesarini, lorenzini, boschi, swinkels - 19:55 Wednesday 28 December 2016 (35933) Print this report
HWS sensor position on EDB
Today we performed some tests on EDB to place the HWS sensor in the image plane of NI. From our test we verified that the angular displacement seen by the sensor is linearly dependent on the amplitude of the NI tilt. We also noticed that the measurement is reproducible only taking into account the automatic alignment of the SDB on B5. We look at the RMS of the gradient amplitudes in the HWS map after a tilt of +/-15 urad of the NI mirror. This measurement has been taken for different positions of the HWS sensor along z axis: the explored range is about 150 mm. Further investigations are needed. We left the HWS sensor in the nominal position of the image plane according to design layout.
AdV-TCS (CO2 laser projector)
cesarini, lorenzini, sposito - 19:40 Wednesday 28 December 2016 (35932) Print this report
Hardware installation for CO2 laser ISS and slow power control
Yesterday we installed all the hardware needed for the implementation of the intensity stabilization servo, except for a photodiode amplifier (that needs to be modified) that will be installed soon, during a lunch break. We also installed all the hardware for the implementation of the slow power control. Both controls will be activated and optimized without interfering with ITF activities.
AdV-TCS (CO2 laser projector)
cesarini, nardecchia, lorenzini - 15:11 Thursday 15 December 2016 (35836) Print this report
Comment to centering WI CH (35703)
The value of the curvature must be divided by 2, due to a double pass measurement.
Search Help
×

Warning

×